Thermal Instruments have been designing and manufacturing primary process control instruments for measuring temperature, flow, pressure, level, and related accessories. Fixed points are the most accurate devices available for defining a temperature scale. Fixed point cells are designed to realize the liquid-solid equilibrium temperatures of certain high purity metal elements, for calibration of thermometers at ITS-90 fixed points.

Master Instruments

Instruments We Calibrate

  • TEMPERATURE INDICATOR / CONTROLLER / SCANNER (WITHOUT SENSOR)
  • TEMPERATURE INDICATOR / CONTROLLER / SCANNER (WITH SENSOR)
  • TEMPERATURE SENSOR
  • TEMPERATURE TRANSMITTER
  • TEMPERATURE RECORDER
  • OVEN / BOD INCUBATOR / FREEZER / REFRIGERATOR
  • FREEZER OR REFRIGERATOR (USING SINGLE SENSOR)
  • OVEN / INCUBATOR / FREEZER (USING NINE SENSORS)
  • MAPPING FURNACE (MUFFLE)
  • TEMPERATURE DATA LOGGER
  • DRY & WET BULB THERMOMETER / SLING PSYCHROMETER
  • TEMPERATURE BATH / DRY-BLOCK CALIBRATOR
  • INFRARED THERMOMETER (NON CONTACT TYPE)
  • TEMPERATURE INDICATOR (SURFACE)
  • TEMPERATURE CUM HUMIDITY INDICATOR (For 10-50°C, 30-90% RH)
  • DEW POINT METER
  • HUMIDITY / ENVIRONMENTAL (STABILITY CHAMBERS) FOR SINGLE
  • HUMIDITY DATA LOGGER
  • HUMIDITY TRANSMITTER
  • HUMIDITY / ENVIRONMENTAL CHAMBERS (9 SENSOR MAPPING)

Our Calibration Scope

S.NoDiscipline / GroupMeasurand or Reference Material / Type of instrument or material to be calibrated or measured / Quantity Measured / InstrumentMeasurement range and additional parameters where applicable(Range and Frequency)Calibration and Measurement Capability(CMC)(±)Calibration or Measurement Method or procedure
1.THERMAL- SPECIFIC HEAT & HUMIDITYINDICATOR of HUMIDITY CHAMBER / GENERATOR / Environmental Chamber @ 25°C30 % RH to 90 % RH2.64 % RHUSING Humidity indicator with sensor by comparison method.
2.THERMAL- SPECIFIC HEAT & HUMIDITYTHERMOHYGROMETER / HUMIDITY SENSOR WITH INDICATOR AND CONTROLLER,TRANSMITTER / DATA LOGGER / RECORDER @ 25°C32.64 % RHUSING DIGITAL HUMIDITY INDICATOR WITH PROBE & HUMIDITY CHAMBER – BY COMPARISON METHOD
3.THERMAL TEMPERATUREEnvironmental chamber / Humidity chamber30 % RH to 90 % RH @25°C4.2 % RHUSING Humidity data logger ( minimum 9 No’s) by mapping method
4.THERMAL TEMPERATUREINFRARED THERMOMETER / PYROMETER / THERMAL IMAGE CAMERA50 °C to 500 °C1.5 °CUSING IR thermometer and BLACK BODY SOURCE – BY COMPARISON
METHOD
5.THERMAL TEMPERATURELIQUID IN GLASS THERMOMETER / Dial temperature gauge(-) 35°C to 250°C0.84°CUSING RTD-PT100 WITH INDICATOR & LIQUID BATH BY COMPARISON
METHOD
6.THERMAL TEMPERATURETEMPERATURE INDICATOR OF LIQUID BATH / FREEZER / OVEN / DRY BLOCK TEMPERATURE CALIBRATOR / INCUBATOR / FURNACE / TEMPERATURE
CHAMBER
(-) 80 °C to 400 °C0.81 °CUSING R THERMOCOUPLE RTD-PT 100 WITH INDICATOR – BY DIRECT METHOD
7.THERMAL TEMPERATURETEMPERATURE INDICATOR OF LIQUID BATH / FREEZER / OVEN / DRY BLOCK TEMPERATURE CALIBRATOR / INCUBATOR / FURNACE / TEMPERATURE CHAMBER400 °C to 1000 °C1.83 °CUSING R THERMOCOUPLE WITH INDICATOR,RTD-PT100 BY DIRECT METHOD
8.THERMAL TEMPERATURETEMPERATURE SENSOR WITH OR WITHOUT INDICATOR
/ CONTROLLER / SCANNER / TRANSMITTER / DATA LOGGER / TEMPERATURE
SWITCH / DIAL THERMOMETER / DIGITAL THERMOMETER / DIAL TEMPERATURE
GAUGE / RECORDER
50°C to 250°C0.6°CUSING DIGITAL TEMPERATURE INDICATOR WITH PROBE & HUMIDITY CHAMBER – BY COMPARISON METHOD
9.THERMAL TEMPERATURETEMPERATURE SENSOR WITH OR WITHOUT INDICATOR / CONTROLLER /
SCANNER / TRANSMITTER / DATA LOGGER /
TEMPERATURE SWITCH / DIAL THERMOMETER / DIGITAL THERMOMETER /
DIAL TEMPERATURE GAUGE / RECORDER
(-) 35 °C to 50 °C0.27 °CUSING DIGITAL TEMPERATURE INDICATOR WITH PROBE & HUMIDITY CHAMBER – BY COMPARISON METHOD
10.THERMAL TEMPERATURETEMPERATURE SENSOR WITH OR WITHOUT INDICATOR / CONTROLLER /
SCANNER / TRANSMITTER / DATA LOGGER /
TEMPERATURE SWITCH / DIAL THERMOMETER / DIGITAL THERMOMETER /
DIAL TEMPERATURE GAUGE / RECORDER
250 °C to 600 °C1.8°CUSING DIGITAL TEMPERATURE INDICATOR WITH PROBE & HUMIDITY CHAMBER – BY COMPARISON METHOD
11.THERMAL TEMPERATURETEMPERATURE SENSOR WITH OR WITHOUT INDICATOR / CONTROLLER /
SCANNER / TRANSMITTER / DATA LOGGER /
TEMPERATURE SWITCH / DIAL THERMOMETER / DIGITAL THERMOMETER /
DIAL TEMPERATURE GAUGE / RECORDER
600°C to 1000°C2.33°CUSING DIGITAL TEMPERATURE INDICATOR WITH PROBE & HUMIDITY CHAMBER – BY COMPARISON METHOD
12.THERMAL TEMPERATURETEMPERATURE SENSOR WITH OR WITHOUT INDICATOR / CONTROLLER /
SCANNER / TRANSMITTER / DATA LOGGER /
TEMPERATURE SWITCH / DIAL THERMOMETER / DIGITAL THERMOMETER /
DIAL TEMPERATURE GAUGE / RECORDER
10 °C to 50 °C0.82 °CUSING DIGITAL TEMPERATURE INDICATOR WITH PROBE & HUMIDITY CHAMBER – BY COMPARISON METHOD